Updating Firmware for FEIG RFID Readers

This page will cover the steps to update firmware for FEIG RFID Readers: CPR74 (HF) and LRU1002 (UHF) . These instructions assume you have a GlueLogix system with a Timing Control Unit (TCU) connected between the PC and the RFID Reader.  If your system is LineLogix Standalone, with no TCU, you can follow the instructions for Connecting to Reader Directly.


Connecting to the Reader Through the TCU:




Connecting to the Reader Directly

If there are any issues with the firmware update, you may wish to connect to the reader directly instead of connecting through the TCU.

Direct Serial Connection:

You can connect to an LRU1002 or a CPR74 RFID Reader with a Serial Connection.

Direct USB Connection:

This section is for LRU1002 Readers only. The CPR74 does not have an easily-accessible USB connection. LRU1002 can connect via Serial or USB, but the USB connection is much faster.


Note: If you cannot connect to your LRU1002 with USB, your reader may not be configured properly. You can either switch to a Serial Connection (with a USB-to-Serial adapter) or reconfigure your reader.
To check USB configuration via the Serial interface, connect ISO Start+ as shown above, and follow the next figure:
See also section UHF: Configuring an LRU1002 Reader in FEIG ISOStart+ Software at https://gluelogix.com/helpFeigSettingsRestore/ConfiguringFEIGReaders.html

You may also need to download the FEIG USB driver here: https://www.feig.de/wp-content/uploads/setup_feigidusb_v3.26.0_win10_build2020120901.zip (from https://www.feig.de/en/drivers/ )

Checking and Updating Reader Firmware












Copyright (c) 2026 GlueLogix Inc., All Rights Reserved.